www.黄色毛片 I 午夜剧场一级片 I 色婷婷中文 I 五级黄色片 I 日本精品一区二区在线观看 I 国产福利三区 I 午夜黄视频 I av无码av高潮av喷吹免费 I 98超碰在线 I 亚洲夜夜爱 I 嘿咻视频在线观看 I 天天爽夜夜爽视频精品 I 久久久久1 I 欧美一级片网址 I 色老二导航 I 久久久一区二区精品 I 你懂的91 I 亚洲乱码中文论理电影 I 国产黄色免费片 I 亚洲国产精品久久久久秋霞影院 I 99中文字幕 I 中文字幕四区 I 在线三级av I 青青草手机视频 I www.亚洲视频.com I 欧美日韩国产成人高清视频 I 国产精品爱久久久久久久电影 I 免费嗨片首页中文字幕 I 天天色综合4 I 日韩中文字幕免费在线观看 I 久久国产精品网站 I 在线免费观看福利影院 I 成人丁香在线 I 米奇777影视狠狠狠 I 色噜噜狠狠色综合欧洲selulu

Products

current position:home page>Products>optical instrument>Ophir>

RM9 7Y70669 with Chopper

Updated:2025-03-13

Views:162

  • Brand: OPHIR
  • Model: RM9
  • Description: RM9 radiometer system is a combination of sensor and beam chopper, which is used to measure the power of very low level CW or quasi CW light source. The RM9 sensor has a pyroelectric element, which ca
  • Feedback
  • WeChat

    QQ

    Online Service

Product Details

Overview of RM9 radiometer system:

User's Manual

RM9 radiometer system is a combination of sensor and beam chopper, which is used to measure the power of very low level CW or quasi CW light source. The RM9 sensor has a pyroelectric element, which can measure radiation in a wide dynamic range, from less than 100nW to 100mW. It has an 8mm aperture and covers the spectral range from 0.15 μ m to 12 μ m. The system has an RMC1 chopper placed between the source and the RM9 sensor.  When installing the chopper, the side marked "THIS SIDE TOWARD SENSOR" should face the sensor.

RM9 Radiometer Drawing:

Main parameters:

Absorber Type Pyroelectric with Chopper
Aperture ?8mm
Spectral Range 0.15-12μm
Power Range 100nW-100mW
Dimensions ?62 W x 21 D (mm)
Max Average Power Density 5W/cm2
Response Time 3.5 s
Max Average Power 100mW

Problem summary:

How to use RM9 radiometer to greatly improve the measurement accuracy?

RM9 is only sensitive to signals chopped at 18 Hz, so placing the chopper as close to the laser source as possible can greatly reduce stray light entering the chopper and being read as part of the signal.

The noise specification is based on a 10 second moving average and the power meter is set to the average of the measured values for better performance.

It is recommended to zero the sensor before use by disconnecting the BNC cable between the RM9 sensor and the chopper or closing the chopper.

Can RM9 be used to measure pulsed laser power?

Yes, but RM9 will measure average power, not energy. In addition, pulse rates below 50 Hz may generate additional noise. A pulse rate close to 18 Hz may cause beat problems.