www.黄色毛片 I 午夜剧场一级片 I 色婷婷中文 I 五级黄色片 I 日本精品一区二区在线观看 I 国产福利三区 I 午夜黄视频 I av无码av高潮av喷吹免费 I 98超碰在线 I 亚洲夜夜爱 I 嘿咻视频在线观看 I 天天爽夜夜爽视频精品 I 久久久久1 I 欧美一级片网址 I 色老二导航 I 久久久一区二区精品 I 你懂的91 I 亚洲乱码中文论理电影 I 国产黄色免费片 I 亚洲国产精品久久久久秋霞影院 I 99中文字幕 I 中文字幕四区 I 在线三级av I 青青草手机视频 I www.亚洲视频.com I 欧美日韩国产成人高清视频 I 国产精品爱久久久久久久电影 I 免费嗨片首页中文字幕 I 天天色综合4 I 日韩中文字幕免费在线观看 I 久久国产精品网站 I 在线免费观看福利影院 I 成人丁香在线 I 米奇777影视狠狠狠 I 色噜噜狠狠色综合欧洲selulu

Products

current position:home page>Products>Measuring Instrument>Campbell>

CR850 Measurement and Control Datalogger

Updated:2025-03-13

Views:172

  • Brand: 美國Campbell
  • Model: CR850
  • Description: The CR850 is a smaller, research-grade data logger that includes an on-board keyboard display as part of its integrated package. It is intended for smaller configurations in which fewer sensors will b
  • Feedback
  • WeChat

    QQ

    Online Service

Product Details

CR850 Measurement and Control Datalogger Description:

Product ManualDatasheet

The CR850 is a smaller, research-grade data logger that includes an on-board keyboard display as part of its integrated package. It is intended for smaller configurations in which fewer sensors will be measured. Each CR850 reads input from sensors, then transmits the data via a communication peripheral; most sensors and telecommunication devices are compatible. Multiple CR850s can be configured as a network or units can be deployed individually. This rugged data logger can provide stand-alone operation in harsh, remote environments.

Detailed Description

The CR850 consists of measurement electronics encased in a plastic shell with an integrated wiring panel, 16- character keyboard, and display screen. The display can show 8 lines x 21 characters (64 x 128 pixels). Custom menus are supported allowing customers to set up choices within the data logger program that can be initiated by a simple “toggle” or “pick list”. Its keyboard and display screen is used to program the data logger, manually initiate data transfer, and display data.

The CR850 uses an external power supply. Low power consumption allows the it to operate for extended periods on a battery recharged with a solar panel—eliminating the need for AC power. The CR850 suspends execution when primary power drops below 9.6 V, reducing the possibility of inaccurate measurements.

The on-board operating system includes measurement, processing, and output instructions for programming the data logger. The programming language, CRBasic, uses a BASIC-like syntax. Measurement instructions specific to bridge configurations, voltage outputs, thermocouples, and pulse/frequency signals are included. Processing instructions support algebraic, statistical, and transcendental functions for on-site processing. Output instructions process data over time and control external devices.

Benefits and Features

Integrated keyboard and display screen let you program, manually initiate data transfers, and view data, all on site.

Ideal applications include wind profiling, weather stations, ETo/agriculture, air quality, soil moisture, water level/stage, aquaculture, vehicle testing, Time Domain Reflectometry, SCADA, and water quality

Simpler in design, the CR800 and CR850 are easier to program and wire.

Serial communications with serial sensors and devices supported via I/O port pairs

Contains custom ASIC chip that expands pulse count, control port, and serial communications capabilities

Supports PakBus, Modbus, SDI-12, and DNP3 protocols

Compatible with channel expansion peripherals allowing you to expand your system

Includes both an CS I/O port and an RS-232 port for connecting communication devices

Gas Discharge Tube (GDT) protected inputs

Battery-backed clock that ensures accurate time is maintained while data logger is disconnected from battery power

Program with LoggerNet, PC400, or Short Cut to fit your setup

CR850 Measurement and Control Datalogger Specifications:

Operating Temperature Range
  • -25° to +50°C (standard)
  • Non-condensing environment
Maximum Scan Rate 100 Hz
Analog Inputs 6 single-ended or 3 differential (individually configured)
Pulse Counters 2
Voltage Excitation Terminals 2 (VX1, VX2)
Communications Ports
  • CS I/O
  • RS-232
Switched 12 Volt 1 terminal
Digital I/O
  • Certain digital ports can be used to count switch closures.
  • 4 I/Os or 2 RS-232 COM
    I/O ports can be paired as transmit and receive for measuring smart serial sensors.
Input Limits ±5 V
Analog Voltage Accuracy ±(0.06% of reading + offset) at 0° to 40°C
ADC 13-bit
Power Requirements 9.6 to 16 Vdc
Real-Time Clock Accuracy ±3 min. per year (Correction via GPS optional.)
Internet Protocols FTP, HTTP, XML POP3, SMTP, Telnet, NTCIP, NTP
Communication Protocols PakBus, Modbus, DNP3, SDI-12, SDM
Idle Current Drain, Average 1 mA (@ 12 Vdc)
Active Current Drain, Average
  • 1 to 16 mA (1 Hz sample rate @ 12 Vdc without RS-232 communication)
  • 16 mA (100 Hz sample rate @ 12 Vdc without RS-232 communication)
  • 28 mA (100 Hz sample rate @ 12 Vdc with RS-232 communication)
Dimensions 24.1 x 10.4 x 5.1 cm (9.5 x 4.1 x 2 in.)
Weight 0.7 kg (1.5 lb)

WeChat

Customer Service QQ

Customer Hotline:

010-52867770
0l0-52867771
l3811111452
l7896OO5796

Technical Supports

0l0-52867774
l7896OO5796
info@dorgean.com